Chapter 6 – Forces 233
parabolic shape. Whereas the local stiffness of the sample is often difficult to determine dir
ectly, the well-defined geometry and mechanical properties of the cantilever enable a more
accurate estimation of its stiffness to be made.
For standard mode force sensing, the AFM cantilever oscillations can be approximated
as those of a solid beam flexing about its point of attachment, namely, that of a beam-like
structure that has one relatively compliant axis (normal to the cantilever flat surface) in terms
of mechanical stiffness, with the two remaining orthogonal axes having comparatively high
stiffness values. The beam bending equation of classical mechanics, for which one end of the
beam is fixed while the other is allowed to undergo deflections, can be used to estimate the
bending stiffness parallel to the z-axis, kz as
(6.20)
k
Yw
w
w
z
y
y
x
=
4
3
where
wi is the width of the cantilever beam parallel to the ith axis (Figure 6.6c)
Y is the Young’s modulus
This beam–cantilever system also has a resonance frequency ν0 given by
(6.21)
v
k
m
w
w
Y
z
z
x
0
0
2
1
2
0 162
=
=
π
ρ
.
where
m0 is the effective total mass of the cantilever and AFM tip
ρ is the cantilever density
Measurement of the resonance frequency far from the surface (such that the potential energy
function is dominated solely by the mechanical elasticity of the cantilever with negligible
contributions from surface-related forces) can be used to estimate kz. However, this relies
on accurate knowledge of the material properties of the cantilever, which are often not easy
to obtain due to the variability from cantilever to cantilever in a given manufactured batch.
An alternative method involves measuring the vertical mean squared displacement (
)
z2
far away from the sample surface, which requires only an accurate method of measuring z.
Deflections of the cantilever can be very accurately detected usually involving focusing a laser
onto the reflecting back of the polished metal cantilever and imaging the reflected image onto
a split photodiode detector. The voltage response is converted ultimately to a corresponding
distance displacement of the tip, provided that the cantilever stiffness has been determined.
We can then use the equipartition theorem in a similar way as for quantifying the stiffness
of optical and magnetic tweezers to estimate k
k T
z
z
B
=
〈
〉
/
2 . At a room temperature of
~20°C, kBT is ~4.1 pN⋅nm (see Chapter 2). Thus, if the voltage output V from a photodiode
results in a volts per nm of cantilever vertical deflection, the z stiffness is given roughly by
4.1a2/〈V2〉 in units of pN/nm. Typical cantilever stiffness values used for probing biological
material in AFM are ~0.1 pN/nm (see Worked Case Example 6.2).
The interaction forces experienced by the tip include electrostatic and chemical forces as
well as van der Waals (vdW) forces. The Morse potential is a good qualitative model for the
chemical potential, in characterizing the potential energy due to the separation, z, of two
atoms that can form a chemical bond to create a diatomic molecule when they approach each
other to within ~0.1 nm, as might occur between atoms of the sample and the approaching
tip. The shape of the potential energy curve has a minimum corresponding to the equilibrium
atom separation, σ, which is ~0.2 nm, so the term potential energy well is appropriate. The
functional form of the Morse potential is
(6.22)
U
E
Morse
bond
exp
(z
)
exp
(z
)]])
= −
−
−
−
[
[ −
−
(2
2
κ
σ
κ
σ